IMPROVEMENT OF AN ANALYTICAL METHOD BASED ON HPLC WITH REFRACTIVE INDEX DETECTION FOR THE ANALYSIS OF GLYCEROL OXIDATION PRODUCTS

  • Cristian J. Giertyas,
  • Débora S. da Silva,
  • Camila L. F. da Silva,
  • Mario R. Meneghetti,
  • Simoni M. Plentz Meneghetti,
  • Rusiene M. de Almeida,
  • Janaína H. Bortoluzzi

DOI
https://doi.org/10.21577/0100-4042.20170391
Journal volume & issue
Vol. 42, no. 7
pp. 729 – 735

Abstract

Read online

The optimized analytical methodology described herein is based on an ion exchange column and an H3PO4 solution (pH 2.24) as the mobile phase, with a flow rate of 0.4 mL min-1 and refractive index detection (RID). On applying the methodology, it was possible to identify 7 products obtained from the oxidation of glycerol with an analysis time of 27 min. The developed method was validated through the evaluation of a series of analytical parameters. The results obtained were evaluated considering the peak area and peak height. The analytical curves showed a correlation coefficient of ≥ 0.986. The coefficient of variation values obtained were ≤ 3.60% for instrumental precision, ≤ 19.36% for intermediate precision in LOQ (1.25 µg mL-1) and ≤ 17.93% for repeatability in LOQ (1.25 µg mL-1) and different days. The limit of quantification established for all compounds was 1.25 µg mL-1 obtained through the parameters of the analytical curve. The accuracy of the method showed recovery values of 85.6 to 112.3% for real fortified sample at 3 concentration levels. Two different samples of glycerol oxidation products were applied to the validated methodology; one obtained from AuNP/SiO2 as catalyst (conversion of 62.42%), and other using AuNP/MWCNT (conversion of 89.5%).